IEEE Transactions on Nuclear Science · 1999 · 11 citations · 8 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic Nuclear MedicineDiagnostic RadiologyAttenuation CorrectionImage Quality AssessmentMedical ImagingNuclear Medicine ImagingPositron Emission TomographyScatter CorrectionAttenuation Correction SystemsContrast-to-noise Ratio EstimationCount Rate EstimationPet Attenuation CorrectionPet QuantificationScatter Fraction EstimationSignal-to-noise RatioWhole-body Pet Imaging
The authors studied the effects of adjusting the low level energy discriminator (LLD) for positron volume imaging (PVI) mode acquisitions using a GE Advance PET system. NEMA scatter fraction and count loss measurements were performed for a 20 cm right circular cylinder at LLD between 300 keV and 450 keV. From these data, noise equivalent count rate (NECR) curves were calculated to estimate effects of LLD on scans of head sized objects. To evaluate the effects of LLD on whole-body image quality, contrast-to-noise ratio (CNR) values were obtained from images processed without attenuation or scatter correction. CNR data were acquired for 1.6 cm and 1.0 cm spherical lesions within the liver region of a Data Spectrum torso phantom scanned with arms, and activity outside the field of view. The results indicate that the NECR for a head-sized object is fairly insensitive to LLD differences between 300 keV and 425 keV for activity concentrations up to 0.4 μCi/cc. The NECR improved slightly for higher concentrations with 375 keV < LLD < 425 keV. Torso phantom images processed without scatter or attenuation correction show some CNR improvement with increasing LLD up to 425 keV for clinically relevant activity concentrations.
8
Performance characteristics of a whole-body PET scanner.
Timothy R. DeGrado, Timothy G. Turkington, John Williams et al. · PubMed · 1994 · 516 citations
Model-based scatter correction for fully 3D PET
John Ollinger · Physics in Medicine and Biology · 1996 · 389 citations
Optimization of noise-equivalent count rates in 3D PET
Ramsey D. Badawi, Paul Marsden, B. F. Cronin et al. · Physics in Medicine and Biology · 1996 · 78 citations