EngineeringCardiologyAlgorithmsBiomedical Signal ProcessingPhysiological Systems ModelingAlgorithm AnalysisBlood Pressure ManagementCardiovascular AssessmentHemodynamicsMethod ComparisonPressure MeasurementAlgorithm ComparisonBlood PressureBlood Pressure EstimationBlood Pressure MeasurementDiastolic Blood Pressure EstimationLinear ApproximationLinear Function ApproximationSystolic Blood Pressure Estimation
Three different algorithms for finding blood pressure through the oscillometric method were researched and assessed. It is shown that these algorithms are based on two different underlying approaches. The estimated values of systolic and diastolic blood pressure are compared against the nurse readings. The best two approaches turned out to be the linear approximation algorithm and the points of rapidly increasing/decreasing slope algorithm. Future work on combining these two algorithms using algorithm fusion is envisaged.