Addiction · 2006 · 12 citations · 21 references
PharmacyPharmacologyToxicologyAlcohol ToxicologyPharmaceutical ComplianceSubstance Use DisordersBioanalytical Method ValidationBreath Alcohol TestingAlcohol Breath AnalysisBreath Alcohol Concentration MeasurementBreath Alcohol MeasurementBreath Analysis MethodologyBreathalyzer InstrumentationComplianceExhaled Breath AssessmentPreliminary Breath Screening
AIMS: To evaluate the ability of a hand-held breath analyser, the Zenalyser((R)) (Zenics Medical), to identify alcohol-dependent patients receiving disulfiram therapy and to assess the sensitivity and specificity of the instrument at different time intervals post-disulfiram dosing. DESIGN: Breath samples were taken from two groups of alcohol-dependent patients, one group on a daily disulfiram regimen and one group receiving no disulfiram. The breath samples were analysed for the combined concentration of carbon disulphide and acetone produced from the metabolism of disulfiram. From these data, two reference ranges were prepared and used for sensitivity and specificity assessments. SETTING: Breath samples for the reference ranges were obtained from patients at Shelton Hospital, Shrewsbury. Breath samples used to assess the sensitivity and specificity of the instrument were obtained from patients at the Edinburgh Alcohol Problems Clinic. PARTICIPANTS: Twenty in-patients from Shelton Hospital receiving a daily 200 mg disulfiram regimen and 20 in-patients receiving no disulfiram. At the Edinburgh Clinic, 54 patients taking a thrice-weekly disulfiram regimen and 22 patients not on disulfiram. MEASUREMENTS: A total of 489 breath samples from Shelton Hospital and 391 breath samples from the Edinburgh Clinic were analysed for the combined concentrations of carbon disulphide and acetone. FINDINGS: The breath analyser produced results that distinguished between the disulfiram-treated and untreated groups (P < 0.001). At 1 day post-dose, the sensitivity was 100% and the specificity was 100%. At 2 and 3 days post-dose, the sensitivities and specificities were 84.6% and 100% and 88.2% and 100%, respectively. CONCLUSION: The breath analyser can improve the assessment of the compliance status of patients receiving a daily dose regimen of disulfiram, but is less useful for this purpose if disulfiram is taken on a thrice-weekly regimen.
21
Disulfiram treatment of alcoholism. A Veterans Administration cooperative study
Richard K. Fuller · JAMA · 1986 · 546 citations
Disulfiram Treatment of Alcoholism
Richard K. Fuller · JAMA · 1986 · 417 citations
Disulfiram Treatment of Alcoholism
Jonathan Chick, Kevin Gough, Wojciech Fałkowski et al. · The British Journal of Psychiatry · 1992 · 369 citations
A DRUG SENSITISING THE ORGANISM TO ETHYL ALCOHOL
Jens Hald, Erik Jacobsen · The Lancet · 1948 · 278 citations