Concepedia

Publication | Closed Access

Downloadable Algorithm to Reduce Inappropriate Shocks Caused by Fractures of Implantable Cardioverter-Defibrillator Leads

140

Citations

13

References

2008

Year

Abstract

A lead-integrity algorithm developed for download into existing implantable cardioverter-defibrillators increases short-term warning of inappropriate shocks in patients with lead fractures and reduces the likelihood of inappropriate shocks. It is the first downloadable RAMware to enhance the performance of nominally functioning implantable cardioverter-defibrillators and the first implantable cardioverter-defibrillator monitoring feature that triggers real-time changes in ventricular fibrillation detection parameters to reduce inappropriate shocks.

References

YearCitations