Flow Injection Analysis of Glucose in Human Serum by Chemiluminescence

33

Citations

8

References

1982

Year

Abstract

Abstract Combining the rigidly controlable conditions of Flow Injection Analysis with detection by chemiluminescence, a method for determination of glucose in serum via enzymatic degradation by glucose in serum via enzymatic degradation by glucose oxidase is described. With a Standard deviation of less than 2% the sampling frequency is 75 samples per hour. Results obtained by the proposed procedure are compared with those acquired by a routine AutoAnalyzer method used at a local hospital. It is shown that any influence on the yield of chemiluminescence due to difference in viscosity of the injected samples effectively can be eliminated.