Activation Time Determination by High‐Resolution Unipolar and Bipolar Extracellular Electrograms in the Canine Heart
82
Citations
22
References
1995
Year
(1) The accuracy of bipolar electrograms is sensitive to wavefront direction, bipole orientation, and interpolar distance; (2) peak negative voltage of uniphasic and triphasic signals is a reliable predictor of AT, but only for B wave; (3) a maximum interpolar distance of 2 mm and bipole orientation parallel to the direction of the impulse wavefront are minimally required for accurate determination of AT during impulse propagation initiated near the recording electrodes; and (4) for impulses initiated near the recording site in normal tissue, a biphasic or triphasic morphology almost certainly indicates that the bipolar electrode is oriented perpendicular to the wavefront direction, irrespective of fiber orientation.