Patent Ductus Arteriosus Equipment and Technique. Amplatzer Duct Occluder: Intermediate‐Term Follow‐Up and Technical Considerations

40

Citations

18

References

2001

Year

Abstract

Between May 1997 and June 2000, 69 patients, ages 0.1 to 34 years, underwent attempted anterograde transcatheter closure of a patent ductus arteriosus (PDA) using the Amplatzer Duct Occluder (ADO). The ADO is a cone-shaped, self-centering, and repositionable occluder made of nitinol wire mesh. A 5Fr to 7Fr sheath was used for the delivery of the device. The mean PDA diameter (at the pulmonary end) was 4.6 +/- 1.9 mm (range 1 mm-8.5 mm). Sixty-seven of the 69 patients had successful device placement. The mean ADO smallest diameter was 6.9 +/- 1.8 mm (range 4 mm-12 mm). Complete angiographic closure occurred in 62 (92.5%) of 67 patients (95% confidence interval, 88.22%-98.77%). In five patients, there was a trivial residual shunt immediately after the procedure. At 24 hours, color Doppler flow imaging revealed complete closure in all 67 (100%) patients. The unsuccessful attempts occurred in two patients with a small, 1-mm diameter native PDA and residual PDA after surgical occlusion. Fluoroscopy time was 7.6 +/- 1.8 minutes (4 min-18 min). No complications were observed. At a median follow-up of 1.5 years (range 0.25 to 3.2 years), all patients had complete closure without complications. We conclude that transcatheter closure using the ADO is a highly effective and safe treatment for most patients with PDA.