Independent component analysis applied to pulse oximetry in the estimation of the arterial oxygen saturation (S<inf>p</inf>O<inf>2</inf>) - a comparative study

T. Jensen, Sune Duun, Jan Larsen, Rasmus G. Haahr, Mette Toft, Bo Belhage, Erik Vilain Thomsen

2009 · 19 citations · 16 references

DOIFull text

Open access

Concepts

Abstract

We examine various independent component analysis (ICA) digital signal processing algorithms for estimating the arterial oxygen saturation (SpO2) as measured by a reflective pulse oximeter. The ICA algorithms examined are FastICA, Maximum Likelihood ICA (ICAML), Molgedey and Schuster ICA (ICAMS), and Mean Field ICA (ICAMF). The signal processing includes pre-processing bandpass filtering to eliminate noise, and post-processing by calculating the SpO2. The algorithms are compared to the commercial state-of-the-art algorithm Discrete Saturation Transform (DST) by Masimo Corporation. It is demonstrated that ICAMS and ICAMF perform up to 13% better than DST. PPG recordings are done with a reflective pulse oximetry sensor integrated in an Electronic Patch. This system is intended for patients with chronic heart and lung conditions.

References

16