MedicineNuclear MedicineRadiologyDiagnostic RadiologyRadiation DosimetryRadiation ProtectionDose CalculationIn Vivo DosimetryPersonnel DosimetryPositron Emission TomographyRadiation Dose MeasurementRadiation Risk AssessmentComputed Tomography Dose MeasurementMedical Imaging Radiation DoseNuclear Medicine Patient DosimetryPatient Dose AssessmentPatient Dose MeasurementRadiation Dose AssessmentRadiation Dose EstimationThermoluminescent Dosimeters
The introduction of PET/CT at King Faisal Specialist Hospital and Research Centre for whole body and brain imaging has become favourable for diagnosis of cancer. There is no data available on the PET/CT dose to staff and members of the public for different activities of (18)F [fluorodeoxyglucose (FDG)] and for longer patient holding time. The study aims to determine and evaluate staff and public doses by using thermoluminescent dosemeters monthly readings for a 7-month deployment period and by using direct measurements of dose rates at 30 cm and 1 m distances from the patients after injection. The whole body doses per procedure and per administered activity of (18)F (FDG) were estimated. A dose map inside the PET/CT was generated to provide information of the dose levels in different locations. The Pearson correlation showed a strong correlation (r(2) = 0.71) between the dose per activity and the number of patients. Optimisation of radiation protection of staff and members of the public was investigated and recommendations were given.