Experimental optimization of the probe for eddy-current displacement transducer

D. Vyroubal, D. Zele

IEEE Transactions on Instrumentation and Measurement · 1993 · 17 citations · 1 references

Concepts

Abstract

The experimental optimization of the coil for an eddy-current displacement transducer probe is presented in the paper. The coil geometry is optimized for the special design of the transducer as a meter of the real part of the probe's impedance. Measurement of the transducer's transfer function has been performed with several samples of ring-shaped coils, as well as with the flat, pancake-shaped coil and with the single-layered cylindrical coil. For the sake of comparison, measurement with the ferrite-cored coil has been included too. The measurement results clearly indicate that the optimal coil geometry is a ring with a uniform and very small cross section of the winding.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

References

1