IEEE Transactions on Biomedical Engineering · 2000 · 47 citations · 12 references
EngineeringBioengineeringBiomedical EngineeringBioinstrumentationElectrophysiologyInstrumentationArtifact RejectionBiomedical InstrumentationBiopotential MeasurementElectromyogram AnalysisMuscle Signal ProcessingCuff ElectrodesElectromyographyElectromyography Artifact RemovalElectromyography InstrumentationElectromyography Signal AcquisitionElectromyography Signal FilteringElectrophysiological RecordingsSource Resistance
A theoretical investigation of different electroneurogram recording techniques using electrode cuffs is presented. A new screened tripole arrangement is proposed with a higher inherent signal to interference ratio than the true tripole, which also allows the nulling of the residual electromyogram signal. The reduction in interference is small because the electrode impedance is large compared to the source resistance.
12
Stable long-term recordings from cat peripheral nerves
R. B. Stein, T. Richard Nichols, Jack H. Jhamandas et al. · Brain Research · 1977 · 161 citations