Test-beam study of the performance of the microstrip gas avalanche chamber

Franco Angelini, R. Bellazzini, A. Brez, M.M. Massai, G. Spandre, Massimo Torquati, R. Bouclier, J. Gaudaen, F. Sauli

IEEE Transactions on Nuclear Science · 1990 · 15 citations · 2 references

DOIFull text

Open access

Concepts

Abstract

The results of a test-beam study of the performance of the microstrip gas avalanche chamber (MISGAC) for the detection of minimum ionizing particles are reported. A detection efficiency of 95%, a position resolution of 30 mu m (RMS), and a two-track resolution of 250 mu m have been measured using a 9-GeV particle beam at the CERN PS. Xenon-DME (dimethyl-ether) and argon-DME mixtures have been used as gas filling for a detector having a total gas thickness of 5 mm.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

References

2