Concepedia

Publication | Closed Access

Elevated Serum Cobalt With Metal-on-metal Articulating Surfaces

183

Citations

45

References

1997

Year

Abstract

We determined serum cobalt levels in 55 patients by atomic absorption spectrophotometry before and after implantation of uncemented total hip arthroplasties. In a randomised, prospective trial 27 wrought Co-28Cr-6Mo-0.2C metal-on-metal articulations were compared with 28 ceramic-on-polyethylene hips which did not contain cobalt. Other sources of iatrogenic cobalt loading were excluded. The metal-on-metal group produced detectable serum cobalt levels (median 1.1 microg/l after one year) which were significantly different (p < 0.0001) from those of the ceramic-on-polyethylene control group (median below detection limit of 0.3 microg/l after one year). Our findings indicate that metal-on-metal bearings generate some systemic release of cobalt.

References

YearCitations

Page 1