Micromotor of Less Than 1 mm^3 Volume for In Vivo Medical Procedures

Brett Watson, James Friend, Leslie Y. Yeo

2009 · 16 citations · 16 references

Concepts

Abstract

The body's stress response to surgery has been cited as a primary cause of post-operative morbidity and has prompted growth in minimally invasive surgical techniques. The future of such techniques lies in the use of in vivo procedures, but is currently limited by the availability of motors with a volume of less than 1 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sup> . In response to this we present a piezoelectric ultrasonic resonant micromotor with a volume of approximately 0.75 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sup> . The motor has a novel helically cut stator that couples axial and torsional resonant frequencies, excited by a lead zirconate titanate element 0.03 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sup> in volume. The motor performance reaches a start-up torque of 47 nNm and no load angular velocity of 830 rad/s. This gives the motor a power density of 18.4 kW/m <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sup> . This performance is on the order necessary to propel a swimming microbot in small human arteries.