Journal of Environmental Quality · 2013 · 26 citations · 19 references
Environmental MonitoringEngineeringPollution MonitoringCoastal WaterHigh FrequencyMarine EnvironmentOcean MonitoringMarine PollutionWater TreatmentAir SamplingWater QualityTidal MarshesWater AnalysisWater MonitoringEnvironmental EngineeringWater PurificationMajor FoulingOptical Remote SensingSitu Ultraviolet-visual Spectrometers
The introduction of portable in situ ultraviolet-visual spectrometers has made possible the collection of water quality parameters at a high frequency in dynamic systems such as tidal marshes. The usefulness of this technology is inhibited by fouling of the instrument's optics. In this study, a spectrometer fitted with manufacturer-recommended compressed air optical cleaning was installed in a brackish marsh to determine if fouling interfered with measurements between bi-weekly servicing. During a 2-wk period, the absorbance measured in air at 220 nm increased from 9 to 549 m, indicating major fouling. An antifouling system was developed that reduced the time of exposure of the optics to stream water and used a pressurized fresh water cleaning. After implementation of the system, the absorbance in air increased to at most 63 m after 2 wk of data collection. The dramatic reduction in fouling will allow quality long-term data to be collected using this technology.
19
John R. Helms, Aron Stubbins, Jason D. Ritchie et al. · Limnology and Oceanography · 2008 · 2.9K citations · Full text
B. A. Pellerin, J. Saraceno, James B. Shanley et al. · Biogeochemistry · 2011 · 275 citations