A Novel Technique for Retrieval of a Drug-Eluting Stent After Catheter Break and Stent Loss

Sunil P. Wani, Seung‐Woon Rha, Ji Young Park, Kanhaiya L. Poddar, Lin Wang, Sureshkumar Ramasamy, Ji Mi Moon, Ji Bak Kim, Sang Ryol Ryu, Seung Yong Shin,

Korean Circulation Journal · 2010 · 11 citations · 7 references

DOIFull text

Open access

Concepts

Abstract

Break of a stent delivery catheter and subsequent stent loss (SL) has been a rare event in the drug-eluting stent (DES) era. We here report a case of successful retrieval of a stent after a break if the delivery catheter and SL from a balloon catheter at a culprit lesion. We finally resolved this situation using a simple balloon technique for both the broken stent catheter inside of the guide catheter and the unexpanded stent in the culprit lesion. Thus balloons are an important weapon in our armamentarium in the cardiac catheterization laboratory for urgent retrieval of a lost stent. Their apt use definitely allowed our patient to avoid undergoing emergency cardiovascular thoracic surgery.