Large Particle Penetration During PM<sub>10</sub>Sampling

William B. Faulkner, Raleigh Allen Smith, John Haglund

Aerosol Science and Technology · 2014 · 11 citations · 8 references

Concepts

Abstract

The objective of the present study was to characterize the&#13;\nperformance of a federal reference method (FRM) PM&#13;\n10&#13;\nsize-&#13;\nselective inlet using analysis methods designed to minimize&#13;\nuncertainty in measured sampling efficiencies for large particles&#13;\nsuch as those most often emitted from agricultural operations.&#13;\nThe performance of an FRM PM&#13;\n10&#13;\ninlet was characterized in a&#13;\nwind tunnel at a wind speed of 8 km/h. Data were also collected&#13;\nfor 20 and 25&#13;\nm&#13;\nm particles at wind speeds of 2 and 24 km/h.&#13;\nResults of the present sampler evaluation compared well with&#13;\nthose of previous studies for a similar inlet near the cutpoint, and&#13;\nthe sampler passed the criteria required for certification as a&#13;\nFRM sampler when tested at 8 km/h. Sampling effectiveness&#13;\nvalues for particles with nominal diameters of 20 and 25&#13;\nm&#13;\nm&#13;\nexceeded 3% for 8 and 24 km/h wind speeds in the present study&#13;\nand were statistically higher than both the “ideal” PM&#13;\n10&#13;\nsampler&#13;\n(as defined in 40 CFR 53) and the ISO (1995) standard definition&#13;\nof thoracic particles (&#13;\np&#13;\n&lt;&#13;\n0.05) for 25&#13;\nm&#13;\nm particles leading to the&#13;\npotential for significant sampling bias relative to the “ideal”&#13;\nPM&#13;\n10&#13;\nsampler when measuring large aerosols.

References

8