Physics in Medicine and Biology · 2013 · 12 citations · 15 references
MedicineRadiation OncologyRadiologyDiagnostic ImagingRadiation DosimetryRadiation ProtectionDetector TechnologyElectron Beam PhysicsImage QualityImage Quality AssuranceRadiation DetectorsRadiation Imaging DetectorsAmorphous SiliconElectron AcceleratorsElectron Beam DiagnosticsElectronic Portal ImagingFlatness MeasurementLow-energy Electron Beams
The feasibility of utilizing an electronic portal imaging device (EPID) for the quality assurance of electron beams was investigated. This work was conducted on a Varian 2100iX machine equipped with an amorphous silicon (aS1000) portal imager. The linearity of the imager pixel response as a function of exposed dose was first confirmed. The short-term reproducibility of the EPID response to electron beams was verified. Low (6 MeV), medium (12 MeV) and high (20 MeV) energies were tested, each along with small (6 × 6 cm(2)), medium (10 × 10 cm(2)) and large (20 × 20 cm(2)) applicators. Acquired EPID images were analyzed using an in-house MATLAB code for radiation field size, penumbra, symmetry and flatness. Field sizes and penumbra values agreed with those from film dosimetry to within 1 mm. Field symmetry and flatness constancies were measured over a period of three weeks. The results indicate that EPID can be used for routine quality assurance of electron beams.