IEEE Transactions on Medical Imaging · 1991 · 618 citations · 28 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsImagingScatter CorrectionCompton Scatter EstimationComputational PhantomsComputed Tomography PhantomsPhotopeak AnalysisPhotopeak Scatter CorrectionScatter Correction AlgorithmsSingle-photon EmissionSolid Phantoms
A new method is proposed to subtract the count of scattered photons from that acquired with a photopeak window at each pixel in each planar image of single-photon emission computed tomography (SPECT). The subtraction is carried out using two sets of data: one set is acquired with a main window centered at photopeak energy and the other is acquired with two subwindows on both sides of the main window. The scattered photons included in the main window are estimated from the counts acquired with the subwindows and then they are subtracted from the count acquired with the main windows. Since the subtraction is performed at each pixel in each planar image, the proposed method has the potential to be more precise than conventional methods. For three different activity distributions in cylinder phantoms, simulation tests gave good agreement between the activity distributions reconstructed from unscattered photons and those from the corrected data.
28
The Fourier reconstruction of a head section
L. A. Shepp, B. F. Logan · IEEE Transactions on Nuclear Science · 1974 · 2.2K citations
Improved SPECT quantification using compensation for scattered photons.
Jaszczak Rj, Greer Kl, Floyd Ce et al. · PubMed · 1984 · 561 citations