Pitfalls and Prevention Strategies for Liquid Chromatography-Tandem Mass Spectrometry in the Selected Reaction– Monitoring Mode for Drug Analysis

104

Citations

22

References

2008

Year

Abstract

To prevent problems such as those we experienced, we recommend the use of stable-isotope internal standards when possible, relative retention times, 2 transitions or more per compound when possible, and acceptable relative abundance ratios between transitions, with an experience-based tolerance of +/-15% for transitions with a relative abundance >10% and with an extension to +/-25% for transitions <10% when the concentration is at the limit of quantification. A powerful general unknown screening procedure can help to confirm suspected interferences. Our results indicate that the specificity of screening procedures is questionable for LC-MS/MS analyses performed in the selected reaction-monitoring mode and involving a large number of compounds with only 1 transition per compound.