IEEE Transactions on Nuclear Science · 1990 · 44 citations · 3 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBody ImagingMedical ImagingPositron Emission TomographyScanning TechniquesBackprojectionDiagnostic Quality MeasurementLimited-angle ReconstructionPet Attenuation CorrectionPet Image ReconstructionPositron Emission Tomography ReconstructionWhole-body Pet Imaging
An investigation of rapid whole-body PET (positron emission tomography) scanning was performed with a view towards its eventual use in screening for metastatic lesions with /sup 18/F-fluoride ion or /sup 18/F-fluorodeoxyglucose. Longitudinal images of the body were formed using three techniques. Simple 2-D projection images were formed by resorting the sinogram data. The images were created for a set of angles covering a full 180 degrees to facilitate viewing in a cine format. The effects of summing over a number of angles per view and filtering to reduce noise were investigated. Limited-angle backprojection images were formed using a small range of angles and no reconstruction filter. The resulting volume was then sliced coronally to give planar images as a function of depth. Coronal images were also obtained after reconstruction with a standard filtered backprojection technique, but with no attenuation correction. All three processes provided surprisingly good images of diagnostic quality, which were obtainable in practical scan times.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
3
ECAT: a new computerized tomographic imaging system for position-emitting radiopharmaceuticals
M. E. Phelps, E.J. Hoffman, Huang Sc et al. · 1977 · 241 citations · Full text