IEEE Transactions on Instrumentation and Measurement · 1983 · 177 citations · 6 references
Natural SciencesElectrical EngineeringMaterials ScienceMaterials CharacterizationMaterials PhysicsMetrologyCalibrationDielectric PropertiesDielectric ResponseDielectricsElectrical CharacterizationElectrical MeasurementDielectric CharacterizationDielectric Constant MeasurementDielectric MeasurementElectrical Instrument CalibrationPermittivity MeasurementProbe Calibration
Routine network analyzer calibration procedures in measurements of the dielectric properties of materials using an open-ended coaxial line probe are frequently inadequate and limit the accuracy of measurements. A calibration method, which makes use of liquids whose properties are well known, is proposed to alleviate this limitation. It is shown that even one liquid used as a standard in place of a matched load greatly improves the accuracy of measurements. Theoretical relationships and experimental results as well as some practical suggestions related to the application of this method are given.
6
M.A. Stuchly, Stanislaw S. Stuchly · IEEE Transactions on Instrumentation and Measurement · 1980 · 637 citations
Natural Sciences, Electrical Engineering, Materials Science +17