Inappropriate Shocks Delivered by an ICD as a Result of Sensed Potentials from a Transcutaneous Electronic Nerve Stimulation Unit

DANIEL M. PHILBIN, Mark Marieb, KESHAVA H. AITHAL, Mark H. Schoenfeld

Pacing and Clinical Electrophysiology · 1998 · 59 citations · 3 references

Concepts

Abstract

Although the potential for transcutaneous electronic nerve stimulation (TENS) units to interfere with appropriate function of cardiac pacemakers is well documented, an adverse interaction between a TENS unit and an ICD has never been reported. We describe a patient in whom a TENS unit created an electrical artifact that was interpreted by a transvenous ICD as ventricular fibrillation, leading to the delivery of inappropriate therapy. TENS units should be used with caution in patients with ICDs.