Evaluation of volume vascularization index and flow index: a phantom study
63
Citations
13
References
2008
Year
Our main finding in a laboratory set-up was a considerable overestimation of moving blood volume using 3D power Doppler ultrasound in a single small tube. The degree of overestimation depends on the spatial resolution and on the settings of the ultrasound instrument. When small vessels are involved in a clinical setting, interpretation of VI should take this overestimation of moving blood volume into account.