2005 · 17 citations · 8 references
EngineeringBiomedical EngineeringCardiologyBiomedical Signal ProcessingElectrophysiologyMobile ComputingBiomedical Signal CompressionCardiac Signal ProcessingData CompressionElectrocardiogram AnalysisElectrocardiographyTelemedicineBit Error RateCompression Performance EvaluationCompression RatioDiagnostic Quality MeasurementDiagnostic Quality MetricsEcg Quality AssessmentEcg Signal CompressionElectrocardiogram Compression
Recently, the set partitioning in hierarchical tree (SPIHT) was shown to be an excellent algorithm for ECG compression. However, how it performs in a cellular phone based wireless environment for telemedicine applications is not known. In this paper, a joint design for SPIHT-based ECG data compression method over a next-generation mobile telecardiology testbed based on the 3G cellular phone standard is proposed and the performance of the testbed for the compressed ECG data segments selected from the MIT-BIH arrhythmia database is evaluated in terms of BER (bit error rate), PRD (percent of root-mean-square difference), compression ratio (CR), transmission time, and diagnostic quality. The simulation results show that during the successful transmission of compressed ECG (when BER is less than 10/sup -5/), a CR of 8:1 provides a 87.5% reduction in total transmission time and a higher CR up to 20 can reduce up to 95% of the required time to transmit the ECG. Furthermore, most characteristics of the received ECG waveform, such as P wave, QRS complex, and T wave, can be reserved with clinically acceptable quality.
8