PharmacyChemistryAnalytical ChemistryCombustionElemental AnalysisCombustion AnalysisDetector SystemsGas ChromatographyInstrumental AnalysisMass AnalyzersMass SpectrometryAutomated Elemental AnalysisGas Chromatography-mass SpectrometryIsotope Ratio AnalysisIsotope Ratio Mass SpectrometryMemory Effects
In Groningen, all organic samples for accelerator mass spectrometry (AMS) are combusted in an automatic Elemental Analyzer, coupled to an Isotope Ratio Mass Spectrometer and Cryogenic Trapping System. The Gas Chromatographic (GC) column, part of the Elemental Analyzer system, appeared to be the main cause for memory effects. Therefore we modified the Elemental Analyzer, such that the trapped CO 2 no longer passed the GC column. Our system modification reduced the memory effect significantly, as shown by lower radiocarbon concentration values for anthracite backgrounds, and a much smaller spread in these values. Our modified system can perform up to 40 combustions unattended in about 6 hr.