Simultaneous mass spectrometry and thermoconductivity measurements of end-tidal xenon concentrations: A comparison

David Gur, John M. Herron, Beth S. Molter, Barbara C. Good, Robert E. Albright, John N. Miller, Burton P. Drayer

Medical Physics · 1984 · 12 citations · 0 references

Concepts

Abstract

Studies in which the rate of local cerebral blood flow is determined by xenon-enhanced computerized tomography require measurement of end-tidal gas. A mass spectrometer and a thermoconductivity detector were compared in measuring the concentration of xenon in end-tidal gas both in humans and in a clinically simulated ("breathing bag") system. The experiments showed that the two instruments provided virtually identical results up to a rate of 18 breaths/min, at which rate recorded concentrations showed marginal degradation with the thermoconductivity detector. Given the absence of a significant difference in the ability of the mass spectrometer and the thermoconductivity detector to measure end-tidal xenon concentration, the thermoconductivity detector appears to be a satisfactory alternative in studies measuring the rate of local cerebral blood flow by enhanced computerized tomography.