Diabetes Care · 2004 · 311 citations · 10 references
EngineeringBioengineeringPhysiologyBiosensorsInstrumentationSensorsBiomedical InstrumentationBiomedical SensorsBlood Glucose MonitoringContinuous Glucose MonitoringGlucose HomeostasisGlycemic MonitoringAccuracyBlood Glucose RegulationContinuous Monitoring SensorsGlucose MonitoringGlucose SensorsProcess CharacterizationSensor AccuracySensor Lag Time
The study introduces continuous glucose–error grid analysis (CG‑EGA) as a method to evaluate continuous glucose‑monitoring sensor accuracy for both glucose values and the direction and rate of glucose fluctuations, demonstrated using TheraSense Freestyle Navigator data. CG‑EGA models glucose monitoring as a time‑dependent process, accounting for location, speed, direction, and interstitial time lags, and comprises point‑error grid analysis (P‑EGA) for value accuracy and rate‑error grid analysis (R‑EGA) for directional accuracy. CG‑EGA showed Navigator accuracy of 73.5% at hypoglycemia, 99% at euglycemia, and 95.4% at hyperglycemia, with hypoglycemia detection failures most common, stable point accuracy across glucose rates, but reduced rate accuracy at high glucose levels, indicating CG‑EGA captures temporal characteristics missed by traditional methods.
OBJECTIVE—The objective of this study was to introduce continuous glucose–error grid analysis (CG-EGA) as a method of evaluating the accuracy of continuous glucose-monitoring sensors in terms of both accurate blood glucose (BG) values and accurate direction and rate of BG fluctuations and to illustrate the application of CG-EGA with data from the TheraSense Freestyle Navigator. RESEARCH DESIGN AND METHODS—We approach the design of CG-EGA from the understanding that continuous glucose sensors (CGSs) allow the observation of BG fluctuations as a process in time. We account for specifics of process characterization (location, speed, and direction) and for biological limitations of the observed processes (time lags associated with interstitial sensors). CG-EGA includes two interacting components: 1) point–error grid analysis (P-EGA) evaluates the sensor’s accuracy in terms of correct presentation of BG values and 2) rate–error grid analysis (R-EGA) assesses the sensor’s ability to capture the direction and rate of BG fluctuations. RESULTS—CG-EGA revealed that the accuracy of the Navigator, measured as a percentage of accurate readings plus benign errors, was significantly different at hypoglycemia (73.5%), euglycemia (99%), and hyperglycemia (95.4%). Failure to detect hypoglycemia was the most common error. The point accuracy of the Navigator was relatively stable over a wide range of BG rates of change, and its rate accuracy decreased significantly at high BG levels. CONCLUSIONS—Traditional self-monitoring of BG device evaluation methods fail to capture the important temporal characteristics of the continuous glucose-monitoring process. CG-EGA addresses this problem, thus providing a comprehensive assessment of sensor accuracy that appears to be a useful adjunct to other CGS performance measures.
10
On Information and Sufficiency
S. Kullback, R. A. Leibler · The Annals of Mathematical Statistics · 1951 · 20K citations · Full text
Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood Glucose
William L. Clarke, Daniel J. Cox, Linda Gonder‐Frederick et al. · Diabetes Care · 1987 · 1.4K citations
Medicine, Endocrinology, Biosensors +16