IEEE Transactions on Magnetics · 2004 · 65 citations · 1 references
EngineeringBiomedical EngineeringElectrical EngineeringBiomedical Signal ProcessingElectromagneticsNeural EngineeringFunctional Electrical StimulationImplant TechnologyImplantable Biomedical DevicesImplantable MicrosystemsNeurostimulationWireless Power TransferBiomedical Wireless Power TransferCoil DesignContactless Power TransferElectrical StimulationImplantable Functional Electrical StimulationImplantable Wireless Power TransferMagnetic FluxPower Output
This paper proposes a new contactless power-signal transmission device using electromagnetic coupling for a functional electrical stimulation system, which consists of two coils. With the primary coil outside and the secondary inside the body, the in-vivo coil composed of a coupling of two parts, one to receive power and one to receive signal, each at a different radio frequency. In this study, our approach is based on a structure to implement a simultaneous power-signal transmission system, which enables power and signal for muscle stimulation to be concurrently transferred from outside the body, while allowing for the separation inside. Moreover, the induced voltage of the magnetic flux on signal transmission, suppressed to 0.15 V, the output signal voltage of 2 V or more was established at a signal frequency of 7.7 MHz, thus leading to stable operation. The study also aimed to reach for higher power output of the secondary coil.
1
Externally powered implantable FES system
Kozue Takahashi, M. Kikuchi, Satoshi Takeuchi et al. · 2002 · 13 citations
Engineering, Biomedical Engineering, Electrical Engineering +17