Concepedia

Publication | Open Access

Midterm experience with implantable cardioverter-defibrillators in children and young adults

34

Citations

28

References

2010

Year

Abstract

The ICD was safe and effective in interrupting malignant arrhythmias in children and adolescents with a high risk of sudden cardiac death. However, the occurrence of lead complications is significant. The incidence of therapies delivered by the device, with appropriate and inappropriate shocks, was high and interfered with the quality of life. The most important reason for inappropriate shocks was T-wave oversensing. Careful programming is mandatory to reduce the inappropriate shocks.