Journal of Clinical Microbiology · 2011 · 117 citations · 15 references
Life SciencesMicrobiologyBacteriologyDiagnostic MicrobiologyMass Spectrometric AnalysisAnaerobic BacteriologyMass SpectrometryPathogen DetectionMaldi-tof IdentificationMaldi-tof Mass SpectrometryMass Spectrometry IdentificationMass Spectrometry-based IdentificationMass Spectrometry-based Microbial IdentificationMatrix-assisted Laser Desorption Ionization Time-of-flight Mass SpectrometryMatrix-assisted Laser Desorption/ionization Time-of-flight Mass SpectrometryTime-of-flight Mass Spectrometry
We compared two matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) systems (Shimadzu/SARAMIS and Bruker) on a collection of consecutive clinically important anaerobic bacteria (n = 290). The Bruker system had more correct identifications to the species level (67.2% versus 49.0%), but also more incorrect identifications (7.9% versus 1.4%). The system databases need to be optimized to increase identification levels. However, MALDI-TOF MS in its present version seems to be a fast and inexpensive method for identification of most clinically important anaerobic bacteria.
15
Piseth Seng, Michel Drancourt, Frédérique Gouriet et al. · Clinical Infectious Diseases · 2009 · 1.9K citations
Suzanne Q. van Veen, Eric C. J. Claas, Ed J. Kuijper · Journal of Clinical Microbiology · 2010 · 656 citations · Full text
Alain Bizzini, Christian Durussel, Jacques Billé et al. · Journal of Clinical Microbiology · 2010 · 600 citations · Full text
Abdessalam Cherkaoui, Jonathan R. Hibbs, Stéphane Emonet et al. · Journal of Clinical Microbiology · 2010 · 522 citations · Full text