PubMed · 2000 · 382 citations · 11 references
DentistryImplant DentistryProsthodonticsImplant ProsthodonticsImplant SurgeryProstheticsAbutment DesignImplant BiomechanicsImplant DesignBiomechanics Of Dental ImplantsBone - Implant Interface MechanicsConnection ModelingDental Implant BiomechanicsDental ImplantsDynamic Loading Of ImplantsLong-term StabilityTapered Abutment Design
This paper presents aa comparison between the 8-degree Morse Taper and the butt joint aa connections between an implant and an abutment. Three-dimensional, non-linear finite element models were created to compare the 2 connection principles under equal conditions. The loading configuration was thereby modeled according to a test setup actually used for the dynamic long-term testing of dental implants as required for regulatory purposes. The results give insight into the mechanics involved in each type of connection and are compared to actual findings with the testing machine. The comparison indicates the superior mechanics of conical abutment connections and helps to explain their significantly better long-term stability in the clinical application.
11
Edwin A. McGlumphy, Deborah A. Mendel, Julie A. Holloway · Dental Clinics of North America · 1998 · 340 citations
Elongation and preload stress in dental implant abutment screws.
James E. Haack, Ronald L. Sakaguchi, Ting Sun et al. · PubMed · 1995 · 319 citations
Dentistry, Implant Dentistry, Mechanical Behavior Of Materials +12