Can first responders achieve and maintain normocapnia when sequentially ventilating with a bag-valve device and two oxygen-driven resuscitators? A controlled clinical trial in 104 patients

Gerrit J. Noordergraaf, P. J. van Dun, B. P. Kramer, M. P. Schors, H. P. Hornman, Wibe A. de Jong, Abraham Noordergraaf

European Journal of Anaesthesiology · 2004 · 20 citations · 9 references

Concepts

Abstract

BACKGROUND AND OBJECTIVE: To evaluate the capability of first responders to achieve and maintain normal ventilation of the lungs of victims employing a bag-valve device and two oxygen-driven resuscitators. METHODS: Prospective, controlled, blinded, single-centre clinical trial using a bag-valve device and one of two FR-300 devices, with 20 cmH2O working pressure, and flows of either 24 or 30 L min(-1). One hundred and four patients were analysed. Induction of anaesthesia followed by ventilation of the lungs with a bag-valve device and an Oxylator in manual and automatic modes performed by a fireman first responder. Each series was repeated for three conditions (anaesthesia; anaesthesia plus muscle relaxation, both with facemask; anaesthesia plus relaxation using an endotracheal tube). RESULTS: Patients age 49 +/- 17 yr; 47% males, 48-132 kg. Normocapnia was achieved and maintained in 66% (bag-valve device), 82% (Oxylator). CONCLUSIONS: The use of an oxygen-driven device improves the ability of first responders to achieve and maintain normocapnia even when distracted. Use of the Oxylators improves performance (P < 0.001) vs. the bag-valve device significantly.

References

9