Clinical and Diagnostic Laboratory Immunology · 1997 · 15 citations · 5 references
Life SciencesCell BiologyImmunologyCellular ImmunologyClinical ImmunologyGood Laboratory PracticeBioanalytical Method ValidationCell CharacterizationClinical AnalysisCytologyFlow CytometryCorrelationFlow Cytometric AnalysisFlow Cytometric Cell AnalysisFlow Cytometric ImmunophenotypingFlow Cytometry ImmunophenotypingMultiparameter Flow Cytometry
The FACSCount flow cytometer absolute CD3+, CD3+ CD4+, and CD3+ CD8+ cell counts measured at a field site hospital laboratory in Thailand were compared to FACScan absolute counts obtained at a nearby research laboratory. Correlation coefficients for 208 samples were > or = 0.95. The FACSCount was accurate, and it was easier and less expensive to operate than the FACScan. Additionally, the FACScan-generated lymphocyte percentage value was accurate for use with the FACScan SimulSET software.