Clinical field site evaluation of the FACSCount for absolute CD3+, CD3+ CD4+, and CD3+ CD8+ cell count determinations in Thailand

Nancy L. Young, Puntawat Ponglertnapakorn, Nathan Shaffer, K Srisak, Thanyanan Chaowanachan, V On-Thern, Chonticha Kittinunvorakoon, A Bunwattanakul, S Suksaweang, Vallerut Pobkeeree,

Clinical and Diagnostic Laboratory Immunology · 1997 · 15 citations · 5 references

Concepts

Abstract

The FACSCount flow cytometer absolute CD3+, CD3+ CD4+, and CD3+ CD8+ cell counts measured at a field site hospital laboratory in Thailand were compared to FACScan absolute counts obtained at a nearby research laboratory. Correlation coefficients for 208 samples were > or = 0.95. The FACSCount was accurate, and it was easier and less expensive to operate than the FACScan. Additionally, the FACScan-generated lymphocyte percentage value was accurate for use with the FACScan SimulSET software.