Concepedia

Publication | Closed Access

Diagnostic reference level of computed tomography (CT) in japan

52

Citations

9

References

2011

Year

Abstract

Optimisation of computed tomography (CT) parameters is important in avoiding excess radiation exposure. The aim of this study is to establish the diagnostic reference levels (DRL) of CT in Japan by using dose-length product (DLP). Datasheets were sent to all hospitals/clinics which had CT scanner(s) in Gunma prefecture. Data were obtained for all patients who underwent CT during a single month (June 2010), and the distributions of DLP were evaluated for eight anatomical regions and five patient age groups. The DRL was defined as the 25th and 75th percentiles of DLP. Datasheets were collected from 80 of 192 hospitals/clinics (26 090 patients). DLP for head CT of paediatric patients tended to be higher in Japan compared with DRLs of paediatric head CTs reported from the EU or Syria. Although this study was performed with limited samples, DLP for adult patients were at comparable levels for all anatomical regions.

References

YearCitations

Page 1