Calibration of the total irradiance monitor

G. M. Lawrence, Greg Kopp, G. J. Rottman, J. W. Harder, T. N. Woods, Hung Loui

Metrologia · 2003 · 20 citations · 8 references

Concepts

Abstract

The Total Irradiance Monitor (TIM) is a total solar irradiance (TSI) instrument designed to achieve a relative standard uncertainty (1 s accuracy) of 100 parts per million (ppm) and a precision and long-term uncertainty of 10 ppm/year. This instrument is one of four on the Solar Radiation and Climate Experiment (SORCE), a NASA/EOS satellite mission scheduled for launch in Nov. 2002. The TIM will report four TSI measurements d aily throughout SORCE’s goal 5-year mission life. Instrument Description Electrical Substitution Radiometers (ESRs) measure incident radiant power. The four ambient-temperature TIM ESRs operate in pairs, with a DSP-controlled servo system continually balancing the temperature of an active ESR to its paired reference ESR. Incident sunlight passing through an open shutter and a precision aperture radiatively heats the active ESR. The replacement electrical heater power required by this ESR to maintain thermal balance as the shutter is cycled determines the incident solar irradiance. The relative standard uncertainty of this primary instrument is 100 ppm (1 s) with 2 ppm noise.

References

8