Development and Clinical Evaluation of an Internally Controlled, Single-Tube Multiplex Real-Time PCR Assay for Detection of <i>Legionella pneumophila</i> and Other <i>Legionella</i> Species

Kate Templeton, Sitha A. Scheltinga, Peter Sillekens, Jantine W. Crielaard, Alje P. van Dam, Herman Goossens, Eric C. J. Claas

Journal of Clinical Microbiology · 2003 · 105 citations · 25 references

DOIFull text

Open access

Concepts

Abstract

A multiplex real-time PCR assay for detection of Legionella pneumophila and Legionella spp. and including an internal control was designed. Legionella species, L. pneumophila, and the internal control were detected simultaneously by probes labeled with 6-carboxy-fluorescein, hexachlorofluorescein, and indodicarbocyanine, respectively. Therefore, no postamplification analysis was required in order to distinguish the targets. The sensitivity of both assays was 2.5 CFU/ml, and from analysis of 10 culture-positive and 74 culture-negative samples from patients investigated for legionellosis, 100% agreement was observed by both assays in comparison to culture. Four additional positives were found by the multiplex real-time PCR assay in the Legionella culture-negative samples.

References

25