Medical Physics · 2005 · 171 citations · 15 references
Veterinary MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsMedical ImagingAdaptive FilteringLine IntegralsLow-dose Computed TomographyLow-dose ReconstructionPenalized LikelihoodPenalized Likelihood EstimationSmoothing
We have developed a sinogram smoothing approach for low-dose computed tomography (CT) that seeks to estimate the line integrals needed for reconstruction from the noisy measurements by maximizing a penalized-likelihood objective function. The maximization is performed by an algorithm derived by use of the separable paraboloidal surrogates framework. The approach overcomes some of the computational limitations of a previously proposed spline-based penalized-likelihood sinogram smoothing approach, and it is found to yield better resolution-variance tradeoffs than this spline-based approach as well an existing adaptive filtering approach. Such sinogram smoothing approaches could be valuable when applied to the low-dose data acquired in CT screening exams, such as those being considered for lung-nodule detection.
15
Early Lung Cancer Action Project: overall design and findings from baseline screening
Claudia I. Henschke, Dorothy I. McCauley, David F. Yankelevitz et al. · The Lancet · 1999 · 2.5K citations · Full text
Mass screening for lung cancer with mobile spiral computed tomography scanner
Shusuke Sone, Shodayu Takashima, Feng Li et al. · The Lancet · 1998 · 942 citations