2002 · 32 citations · 3 references
EngineeringBiomedical EngineeringPhysiologyBioinstrumentationBiomedical OpticsBiomedical Signal ProcessingHemodynamicsMicrocirculatory PhysiologyMicrocirculatory TransportBlood FlowBlood Flow MeasurementMicrospheresPressure Control Valves
The authors report on the development and application of automated spectrophotometry for regional blood flow measurement using the coloured microsphere (CM) dye-extraction technique. A software program running under Microsoft Windows 3.1 was written to simultaneously control an autosampler, sipper pump, pressure valves and spectrophotometer with an exclusively designed small-volume flow-cell. The system allowed consecutive measurements of over 200 tissue samples in volumes as small as 40 /spl mu/L without any attendance. The system was validated in eight chronically instrumented dogs. A comparison between the advanced CM technique and the standard radionuclide microspheres method demonstrated a strong correlation for blood flows to various organs in awake animals. The system proved to be a reliable, user-friendly and time-saving application for the CM dye-extraction technique.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
3
Measurement of regional myocardial blood flow with multiple colored microspheres.
PETER KOWALLIK, Rainer Schulz, Brian D. Guth et al. · Circulation · 1991 · 379 citations · Full text