IEEE Transactions on Magnetics · 1991 · 47 citations · 3 references
EngineeringBioengineeringBiomedical EngineeringElectromagneticsElectromagnetismSensorsIron MetabolismMagnetic CharacterizationMagnetic SensorsMagnetic SusceptibilityIron Stores AssessmentLiver Iron Concentration AssessmentLiver Iron Concentration MeasurementMagnetometer InstrumentationMagnetometer MeasurementsMagnetometersSquid MagnetometersSystem Noise Reduction
A novel SQUID magnetometer for noninvasive measurements of hepatic (liver) iron stores is discussed. Placement of the SQUID, detection coil, and magnet in the dewar vacuum region significantly reduced system noise. In addition, the system incorporates multiple magnets and detection coils which may allow the discrimination of the surface skin layer from the deeper (weaker signal) true liver iron concentration. Measurements indicate an instrumental noise level that is <20 mu g/g of equivalent iron concentration.
3
Magnetic-Susceptibility Measurement of Human Iron Stores
Gary M. Brittenham, David E. Farrell, John William Harris et al. · New England Journal of Medicine · 1982 · 372 citations
Medicine, Biophysics, Hematology +17
Magnetic measurement of human iron stores
D. E. Farrell, J. H. Tripp, P.E. Zanzucchi et al. · IEEE Transactions on Magnetics · 1980 · 51 citations
A clinical system for accurate assessment of tissue iron concentration
D. E. Farrell, J. H. Tripp, Gary M. Brittenham et al. · Il Nuovo Cimento D · 1983 · 15 citations