Determination of conjugated bile acids in human bile and duodenal fluid by reverse-phase high-performance liquid chromatography.

C A Bloch, John B. Watkins

Journal of Lipid Research · 1978 · 97 citations · 16 references

DOIFull text

Open access

Concepts

Abstract

A simple mehtod using reverse-phase liquid chromatography is presented for resolution and quantitation of the major conjugated bile acids of man, including the glycine and taurine conjugates of the dihydroxy bile acids, chenodeoxycholic and deoxycholic acid. Using modern, high-performance chromatographic equipment, analysis time is less than 30 minutes. The quantitative range of the method, with detection by refractive index, is 0.05 to 0.1 mumol of bile acid and the limit of detection for an injection sample is 0.01 mumol. This provides a sensitivity sufficient for analysis of dilute duodenal and gallbladder bile with minimal sample preparation.

References

16