IEEE Transactions on Nuclear Science · 1997 · 19 citations · 19 references
Natural SciencesRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyDetector TechnologyGamma-ray DetectorsGamma-ray InstrumentationGamma-ray SourcesRadiation Imaging DetectorsRadiographic ImagingCompton ImagingDetector Spatial ResolutionGamma-ray Detector TechnologyGamma-ray ImagingImaging ResolutionIntrinsic Energy ResolutionReadout ElectronicsSpatial Resolution Measurement
A high pressure xenon scintillation drift chamber with 30 cm diameter field of view has been built and tested as a two-dimensional imaging camera for low energy /spl gamma/-rays. Experiments demonstrate that the camera can be also operated in an electronically collimated mode as a three-dimensional imaging Compton camera. A spatial resolution of 25 mm is measured for a 140 keV point-like source placed 10.5 cm in front of the camera without any collimator. The resolution is limited by the intrinsic resolution of the 19 channel readout system which is based on 80 mm diameter phototubes. A way to improve the imaging performance is discussed.
19
Hal O. Anger · Review of Scientific Instruments · 1958 · 791 citations
R. W. Todd, J.M. Nightingale, Delora Everett · Nature · 1974 · 331 citations
Semiconductor counters for nuclear radiations
Nuclear Physics · 1964 · 180 citations