Artificial Organs · 2009 · 34 citations · 6 references
EngineeringBiomedical EngineeringElectrical EngineeringAssistive TechnologyEnergy SystemsHeat TransferImplant TechnologyImplantable Biomedical DevicesBiomedical Wireless Power TransferBiventricular Assist DevicesEnergyFerrite CoresHybrid Energy SystemsImplantable Wireless Power TransferTransmission Efficiency
We have developed a new transcutaneous energy transmission (TET) system for a totally implantable biventricular assist device (BVAD) system in the New Energy and Industrial Development Organization (NEDO) artificial heart project. The TET system mainly consists of an energy transmitter, a hybrid energy coil unit, an energy receiver, an internal battery system, and an optical telemetry system. The hybrid energy coil unit consists of an air-core energy transmission coil and an energy-receiving coil having a ferrite core. Internal units of the TET system are encapsulated in a titanium alloy casing, which has a size of 111 mm in width, 73 mm in length, and 25 mm in height. In in vitro experiments, the TET system can transmit a maximum electric energy of 60 Watts, and it has a maximum transmission efficiency of 87.3%. A maximum surface temperature of 46.1 degrees C was measured at the ferrite core of the energy-receiving coil during an energy transmission of 20 Watts in air. The long-term performance test shows that the TET system has been able to operate stably for over 4 years with a decrease of energy-transmission efficiency from 85% to 80%. In conclusion, the TET system with the hybrid energy coil can overcome the drawback of previously reported TET systems, and it promises to be the highest performance TET system in the world.
6
The AbioCor implantable replacement heart
Robert D. Dowling, Laman A. Gray, Steven W. Etoch et al. · The Annals of Thoracic Surgery · 2003 · 102 citations
Medicine, Cardiology, Surgery +12
Eiji Okamoto, Yoshiro Yamamoto, Yusuke Inoue et al. · Journal of Artificial Organs · 2005 · 32 citations