Temperature and pressure effects during erbium laser stapedotomy
63
Citations
14
References
1996
Year
From the comparison of the laser-induced pressure with the limit graph to avoid hearing defects published by Pfander, an unobjectionable use of the erbium laser is deduced for fluences < 10 J/cm2. The erbium laser seems to represent an ideal instrument for middle ear surgery with all the advantages (precision, fiber optic transportable, high ablation efficiency, safety) desired for clinical application.