Worldwide proficiency test for routine analysis of <i>δ</i> <sup>2</sup> H and <i>δ</i> <sup>18</sup> O in water by isotope‐ratio mass spectrometry and laser absorption spectroscopy

70

Citations

17

References

2012

Year

Abstract

The leading cause of outliers appeared to be improper calibration or compromised storage of laboratory standard and primary reference waters; hence the importance of judicious storage of lab standards cannot be understated. The deprecated practice of single standard normalization was identified as a problem for some laboratories. We further recommend that laboratories strive to report parsimonious long-term precisions based upon control standards, and to improve quantification and correction for LAS instrumental drift and inter-sample carryover effects.