Repeatability of T1‐quantification in dGEMRIC for three different acquisition techniques: Two‐dimensional inversion recovery, three‐dimensional look locker, and three‐dimensional variable flip angle
39
Citations
18
References
2010
Year
Both 2D-IR and 3D-LL perform well in generating repeatable dGEMRIC results, while 3D-VFA results are somewhat inferior. Furthermore, repeatability results in this study are similar to previously published results for healthy subjects. Finally, the positioning of the analyzed images is crucial to generate reliable repeatability results.