Performance of a Polymer-Based DNA Chip Platform in Detection and Genotyping of Human Papillomavirus in Clinical Samples

Thomas Schenk, Thomas Brandstëtter, Axel zur Hausen, Juliane Alt‐Mörbe, Daniela Huzly, Jürgen Rühe

Journal of Clinical Microbiology · 2009 · 12 citations · 30 references

DOIFull text

Open access

Concepts

Abstract

Human papillomavirus (HPV) plays a key role in the development of cervical and laryngeal cancers. The aim of our study was to compare the performance of a new hydrogel-based HPV genotyping biochip assay (Biochip) to a commercially available and CE-marked conventional PCR followed by reverse hybridization (GenID-PCR). One hundred twenty-three samples were available for the study. Of these samples, 101/123 were gynecological swabs, 8/123 were swabs or biopsy samples of genital warts, 7/123 were biopsy samples of otorhinolaryngeal lesions, 5/123 were samples of skin warts, and 2/123 were samples of orolabial abnormalities. These molecular methods for HPV genotyping showed comparable sensitivity and specificity. However, 19/123 of the results were discrepant. Specifically, Biochip showed better performance in the detection of multiple infections, especially when more than one high-risk genotype was present. Due to the different probe configurations used in the two assays, GenID-PCR achieves only group-specific detection of many HPV genotypes, whereas Biochip allows for specific identification. Overall, the newly developed HPV chip system (Biochip) proved to be a suitable tool for HPV detection and genotyping; it also proved to be superior for establishing HPV genotyping methods.

References

30