Sensors · 2012 · 61 citations · 52 references
Natural SciencesBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsBiophotonicsCellular ImagingFluorescence ImagingFluorescence MicroscopyFluorescence SpectroscopyMolecular Imaging In OncologyTumor ImagingContrastFluorescence LifetimeFluorescence Lifetime ImagingFluorescence Lifetime Imaging MicroscopyHardware InstrumentationIn Vivo Fluorescence ImagingSingle-photon Avalanche DiodesTwo-photon Fluorescence Imaging
We have successfully demonstrated video-rate CMOS single-photon avalanche diode (SPAD)-based cameras for fluorescence lifetime imaging microscopy (FLIM) by applying innovative FLIM algorithms. We also review and compare several time-domain techniques and solid-state FLIM systems, and adapt the proposed algorithms for massive CMOS SPAD-based arrays and hardware implementations. The theoretical error equations are derived and their performances are demonstrated on the data obtained from 0.13 μm CMOS SPAD arrays and the multiple-decay data obtained from scanning PMT systems. In vivo two photon fluorescence lifetime imaging data of FITC-albumin labeled vasculature of a P22 rat carcinosarcoma (BD9 rat window chamber) are used to test how different algorithms perform on bi-decay data. The proposed techniques are capable of producing lifetime images with enough contrast.
52
A transparent access chamber for the rat dorsal skin fold
Heinz D. Papenfuss, Joseph F. Gross, Marcos Intaglietta et al. · Microvascular Research · 1979 · 284 citations