2003 · 10 citations · 2 references
MedicineNeurologyNeuroscienceElectrophysiologyNeural EngineeringNeurophysiologyElectroencephalogram AnalysisElectroencephalogram Signal ProcessingElectroencephalographic MonitoringElectroencephalographyHypoxic-ischemic Brain InjuryNeurologic MonitoringBurst DetectionCerebral Hypoxic-ischemic InjuryEeg Background MonitoringFalse Alarm Rate AnalysisHypoxic-ischemic Brain Injury MonitoringHypoxic-ischemic Brain Injury PrognosisHypoxic-ischemic Encephalopathy MonitoringNeurological Outcome Prediction
Prior research has shown that bursting in early recovery, is an indicator of neurological outcome following a hypoxic-ischemic injury. This paper compares three burst-counting procedures, the "gold standard" clinician, the Teager Energy Operator (TEO) and the Square Law Detector (SLD). Analysis of detection performance showed that the TEO easily detects more accurate bursts than the traditional SLD. During periods of high EEG background activity, the TEO maintains a high burst detection rate coupled with a low false alarm rate.
2
Detecting EEG bursts after hypoxic-ischemic injury using energy operators
David Sherman, Ansgar M. Brambrink, D. Walterspacher et al. · 2002 · 17 citations
Medicine, Neurology, Neuroscience +17