A transconductance driven-right-leg circuit

Enrique Mario Spinelli, Nahuel Martínez, Miguel Mayosky

IEEE Transactions on Biomedical Engineering · 1999 · 75 citations · 6 references

Concepts

TL;DR

Biopotential measurements are highly susceptible to electromagnetic interference, which can enter the acquisition system in differential and common‑mode forms; driven‑right‑leg circuits are widely employed to mitigate common‑mode interference. This paper reports an improvement on the classic DRL. The proposed circuit employs a transconductance amplifier to drive the patient’s body, extending the bandwidth for high‑frequency EMI rejection such as fluorescent‑light interference, and a comparative analysis against a typical DRL is presented. The improvement yields about 20 dB better rejection at a few kilohertz and the circuit is easy to compensate for stability, as confirmed by experimental tests.

Abstract

Biopotential measurements are very sensitive to electromagnetic interference (EMI). EMI gets into the acquisition system by many ways, both as differential and common mode signals, driven-right-leg circuits (DRL) are widely used to reduce common mode interference. This paper reports an improvement on the classic DRL. The proposed circuit uses a transconductance amplifier to drive the patient's body. This configuration has some interesting properties, which provide an extended bandwidth for high-frequency EMI rejection (such as fluorescent lights interference). The improvement is around 20 dB for frequencies of few kilohertz and the circuit is easy to compensate for stability. A comparative analysis against a typical DRL is presented, the results obtained have been experimentally tested.